1.3 Names and Functions of Parts
16
Integration (INTEGRATOR) status indicator lamps (p.62)
Hold (HOLD) status indicator lamp (p.108)
RUN
Indicates the status of integration based on START/STOP key
operation or communications.
RUN lamp on: Integration active
RUN lamp flashing: Integration stopped
RUN lamp off: Integration reset
RUN EXT
Indicates the status of integration based on external control.
RUN lamp on, EXT lamp on: Integration active
RUN lamp flashing, EXT lamp on: Integration stopped
RUN lamp and EXT lamp off: Integration reset
HOLD
Lights up when display hold is activated by pressing the HOLD
key.
To cancel display hold: Pressing the HOLD key once more
causes the display hold state to be canceled and the HOLD lamp
to be turned off.
MAX
When the HOLD, MAX, and MIN lamps are all off, pressing the
SHIFT key and then the HOLD key causes the maximum value
to be held and the MAX lamp to light up.
MIN
Pressing the HOLD key while the MAX lamp is lit up (indicating
that the maximum value is being held) causes the minimum val-
ue to be held and the MIN lamp to light up.
Pressing the HOLD key while the MIN lamp is lit up (indicating
that the minimum value is being held) causes the minimum value
hold to be canceled, returning to the normal measured value dis-
play.