2.2 Connecting the Measurement Lines
30
Connecting the measurement lines
Please read "Handling the Instrument" (p.6) carefully before connecting the instru-
ment to the measurement lines.
Connect cables to the instrument’s voltage and current input terminals.
You will need: A Phillips head screwdriver with a No. 3 tip
Verify that power to the measurement lines has been cut off before connecting the
instrument to them.
To avoid electric shock or a short-circuit at the input terminals, use solderless termi-
nals that cover wiring with insulation.
To prevent instrument damage or electric shock, use only the screws (M6
×12 mm)
for securing the voltage input terminals and current input terminals in place and the
screws (M3
×6 mm) for securing the safety cover in place that shipped with the prod-
uct. If you loose any screw or find that any screws are damaged, please contact your
Hioki distributor for a replacement.
Safety covers
The safety covers play a protective role by preventing contact with the terminals. Always attach the covers
before using the instrument.
Verify that no voltage is being applied to the measurement lines before attaching or removing the safety
covers.
1
Remove the safety covers from the voltage input terminals and current input terminals.
2
Connect the cables to the voltage input terminals and current input terminals.
3
Attach the safety covers.
Remove the safety cover.
2
Remove the screws. (M3 × 6 mm)
1