2.2 Connecting the Measurement Lines
33
Selecting the connection method (Example: 1P2W connection)
Depending on the input level, the power meter’s instrument loss may affect measured values. Select the con-
nection method below with the lower loss.
Example:
When measuring 12 V, 65 A
Method 1: Loss = (12 [V])
2
/ 2 [MΩ] = 0.000072 [W]
Method 2: Loss = (65 [A])
2
× 1 [mΩ] = 4.225 [W]
Method 1 is characterized by lower loss and therefore allows more accurate measurement.
When measuring 1,000 V, 10 mA
Method 1: Loss = (1,000 [V])
2
/ 2 [MΩ] = 0.5 [W]
Method 2: Loss = (10 [mA])
2
× 1 [mΩ] = 0.0000001 [W]
Method 2 is characterized by lower loss and therefore allows more accurate measurement.