3.2 Configuring Settings
43
Example: When the PW3337 wiring mode is 1P2W×3
1 Press .
2 Press to display the settings
screen shown to the left.
:
The screen varies depending on whether the wiring mode is
1P2W×2 or a mode other than 1P2W.
:
The screen varies depending on whether the wiring mode is
1P2W×3, 1P3W&1P2W and 3P3W&1P2W, 3P3W2M, or 3V3A
and 3P3W3M and 3P4W.
3 Press or to select the channel
to set.
Areas b, c, and d on the display correspond to CH1, CH2,
and CH3, respectively. The selected channel parameter will
flash.
4 Press or to set the current
input method.
Settings: OFF (direct input) TYPE1 (BNC
terminal direct connection [using
external sensors]) TYPE2 (con-
nection via the 9555-10 and L9217
[using external sensors])
(When set to TYPE1 or TYPE2, the EXT.SENSOR lamp for
the set channel will light up.)
5 Set the current input method for other
channels as necessary.
When using a wiring mode other than 1P2W, the current input method will be standardized using the CH1
settings.
The current input method cannot be changed while integration is being performed or during display hold or
maximum value/minimum value hold operation.
6 Press to exit the settings.
The instrument will return to the normal measure-
ment state.