112 Manufacturer’s declaration on electromagnetic compatibility
Guidance and manufacturer‘s declaration – electromagnetic immunity
The point 2 is intended for use in the electromagnetic environment speci ed below. The user of the
point 2 should assure that it is used in such an environment.
Immunity test IEC 60601
Test level
Compliance level Electromagnetic environ-
ment – guidance
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
± 6 kV
contact
± 8 kV
air
± 6 kV
contact
± 8 kV
air
Floors should be made of
wood, concrete or ceramic
tile. If oors are covered with
synthetic material, the rela-
tive humidity should be at
least 30%.
Electrical fast transient
(Burst) IEC 61000-4-4
± 2 kV
power supply lines
± 1 kV
input / output
± 2 kV
power supply lines
± 1 kV
input / output
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Surges
IEC 61000-4-5
± 1 kV
differential mode
± 2 kV
common mode
± 1 kV
differential mode
not applicable
Mains power quality should
be that of a typical commer-
cial or hospital environment.