EasyManua.ls Logo

Hoffrichter CPAP - Parameter Settings

Hoffrichter CPAP
120 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Device functions 81
PARAMETER SETTINGS
Parameter Settings range Dependency Setting steps
Language
DEU, ENG, ELL, SPA, FRA, ITA,
NLD, TUR, PLK, POR, CZE
--
P-Unit hPa, mbar, cmH
2
O--
Press.
1
4 – 20 hPa - 0.5 hPa
IPAP
2
EPAP – 20 hPa EPAP, 20 hPa 0.5 hPa
EPAP
2
4 hPa – IPAP 4 hPa, IPAP 0.5 hPa
I-Slope
2
1 – 3
IPAP, EPAP,
Frequency, I:E
1 Level
E-Slope
2
1 – 3
IPAP, EPAP,
Frequency, I:E
1 Level
Frequency
3
5 – 30 bpm - 1 bpm
I:E
3
20 – 80 % - 1%
Delay Time
4
0 – 20 s - 1 s
Backup freq
5
ON, OFF - -
P-Min
6
4 – 20 hPa P-Start 0.5 hPa
P-Max
6
4 – 20 hPa P-Start 0.5 hPa
P-Start
6
4 – 20 hPa P-Min,
P-Max
0.5 hPa
I-FLEX 0 3 hPa/ (l/s) - 1 hPa/ (l/s)
E-FLEX 0 3 hPa/ (l/s) - 1 hPa/ (l/s)

Table of Contents

Related product manuals