62 Device functions
2. Use the selection keys and to set the
desired value. Confi rm your setting by pressing
the enter key .
The parameter E-Slope is set next.
Calculation of the I-Slope time:
Settings:
IPAP = 10 hPa I-Slope = 2 Frequency = 12 bpm
EPAP = 6 hPa E-Slope = 2
I-Slope time =
IPAP – EPAP
I-Slope
I-Slope time =
10 hPa – 6 hPa
20 hPa/s
= 0.2 s
E-SLOPE
(only point 2 BILEVEL ST20 in modes BILEVEL S, T
and ST)
The E-Slope determines the time of pressure decrease
from IPAP to EPAP.
• Level 1 = 10 hPa/s
• Level 2 = 20 hPa/s
• Level 3 = 40 hPa/s