Device functions 71
CALIBRATION
Calibration guarantees optimum performance of
the automatic function and mask test. It must be
performed when:
the device is used initially,
the mask has changed,
there is a change from operation without a
bacterial lter to with a bacterial lter, or vice
versa, or
there is a change from operation without a
humidi er to with a humidi er, or vice versa.
Calibration is only possible during mains opera-
tion and not in battery mode.
Calibrating the device
1. Connect the mask, the therapy tube and, if
applicable, the humidi er or bacterial lter to the
device.
Ensure that the air and exhalation openings in
the mask are clear and the tube is not kinked.
Do not put the mask on.