72 Device functions
2. Press the enter key for 1 second.
3. Use the selection keys
and to select
Calibration.
4. Press the enter key
.
5.
Start appears on the display. Press the enter
key
. If you do not press the key within 5 s,
the display jumps back (see point 3).
6. The blower starts and calibration begins.
Calibration active appears on the display.
Calibration takes 20 - 70 s. The blower switches
off after calibration is nished.
7. If calibration has been successful, Successful
appears on the display and if it has failed,
Failed appears on the display.
8. In both cases press the enter key .
The old values will be retained if calibration
has failed. Start calibration again.