Device functions 61
EPAP
(only point 2 BILEVEL ST20 in modes BILEVEL S, T
and ST)
Selection of expiration pressure (EPAP) by the
physician
1. First set IPAP.
2. Use the selection keys and to set the
desired value. Con rm your setting by pressing
the enter key .
The parameter I-Slope is set next.
I-SLOPE
(only point 2 BILEVEL ST20 in modes BILEVEL S, T
and ST)
The I-Slope determines the time of pressure increase
from EPAP to IPAP.
Level 1 = 10 hPa/s
Level 2 = 20 hPa/s
Level 3 = 40 hPa/s
Selection of I-Slope by the physician
1. First set IPAP and EPAP