A-5
Carm Unable to position device within specified tolerance
Hardware Comment: There is a problem with the aperture mechanism.
Action: Check the aperture assembly. Test the aperture positioning with the SUSQ
program.
Corrupted Scan Data Detected Software This message should be reported to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
Could not find file extension in Software This message should be reported to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
Couldn't write the new record Hardware
or
Software
Comment: The system was unable to write a reference curve that was entered or edited.
Action: Check that the hard drive is not full. That is, check the message "Room for __
scans" in the status window and ensure that it does not say zero. If it is full, then
archive and delete one or more scans and try again.
If the hard drive is not full, then you have either a hard drive failure or a corrupt
reference curve database.
db_File error __ .....
or
db_VISTA error __ .....
Software Any message that begins this way should be reported to Hologic Technical Support.
Include as much information as possible about what you were doing at the time.
Device already in use by foreground task [E_LOCKOUT_I]
MAPI Comment: The MAPI driver was already opened by another application when the
driver attempted to go ON_LINE (i.e., the driver requires exclusive access to the
MAPI driver).
Action: Report this error to Hologic Technical Support.
ERROR MESSAGE
HARDWARE
/SOFTWARE
ERROR
LOCUS COMMENT/ACTION