20 201T User Manual 511391
Honeywell 2/07
Specifications
Specifications
Periodic Inspections and Calibration
This unit requires calibration. The calibration frequency will be a
function of the operating conditions, including operating under extreme
temperatures, exposure to contaminants or gas concentrations greater
than the lower explosive limits. A calibration inspection must be
included as part of a routine maintenance to ensure proper operation of
the gas detection unit.
If unit span or zero cannot be adjusted, the sensor may be approaching
its end-of-life or has been contaminated and must be replaced.
Power Requirements: 17 - 27 Vac, 24 - 38 Vdc, 250 mA
Operating Temperature
Range:
Q1 and Q2 CO : -4°F to 122°F (-20°C to 50°C)
Q2 : 32°F to 100°F (0 °C to 40 °C)
Operating Humidity
Range:
15 to 90% RH, (non-condensing)
Outputs: DPDT optional relay
5A, 30 Vdc or 250 Vac (resistive load)
4-20 mA (optional)
Audible Alarm: 65 dBa - 1 m (option)
Dimensions: 8.4 in(H) x 5.3 in( W ) x 2.25 in(D)
(21.3 cm x 13.4 cm x 5.7 cm )
Certifications
CAN/CSA C22.2
ANSI/UL 61010-1