14 MagneW 3000
PLUS
Specification and Application Guide 1/98
Installation Planning Considerations
The MagneW 3000
PLUS
planning considerations are summarized in Table 5. To ensure proper flowmeter selection
and operation, the following installation conditions should be reviewed:
Environmental conditions
Fluid to be measured
Measured liquid flow conditions
Detector location in piping
Clearance for maintenance and inspection
Table 5—Summary of Installation Considerations
Factor Consideration
Environmental conditions The following should be considered when installing the MagneW 3000
PLUS
.
The ambient temperature should be within the following ranges:
Integral model:
–25 to +60°C (–13 to +140°F)
Remote converter:
–25 to +60°C (–13 to +140°F)
Remote detector:
PFA lining:
–30 to +80°C (–22 to +176°F)
Polyurethane rubber or Chloroprene rubber lining:
–30 to +60°C (–22 to +140°F)
Whenever possible, the flowmeter should not be exposed to direct sunlight, rain, or other
unfavorable weather.
The flowmeter should be installed as far from any pump in the line as practical, so that the flo
w
does not pulsate.
The flowmeter must not be subjected to severe vibration, as equipment damage could result.
The flowmeter must not be subjected to a highly corrosive atmosphere, as equipment damag
e
could result.
The flowmeter must be installed sufficiently apart from high-current power lines, motors,
transformers, or any other source of electromagnetic interference.