8. If you are done performing benzene measurements or if you want to use the MultiRAE or
MultiRAE Pro for TVOC operation, either remove the RAE-Sep Tube Cartridge or press
the button so that the “B” is hidden and the “T” side is fully exposed (all the way out):
9. Press [N/-] to advance to the main reading screen.
IMPORTANT!
Do not overtighten any portion of the sampling assembly.
Note: When the instrument is used only for TVOC (total VOC) monitoring, there is no need to
use a RAE-Sep Tube Cartridge. If the cartridge is left on during VOC monitoring, make sure it is
set for TVOC measurement (the “T” is showing).
Honeywell MultiRAE Series™ 65 User Manual