Honeywell STT850 Series - Table 12 - Tamper Reporting Logic Implementation with Write Protect

Honeywell STT850 Series
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 52 STT850 Series HART/DE Option User’s Manual Revision 4.0
Online/Device
Setup/Display Setup
(Continued …)
(for Basic Display
Only)
Common Setup
The following screen options are available:
Rotation Time: Set the length of time the PV
display is visible before rotating to another
screen (such as a diagnostic display). Select
from 3 to 30 seconds.
Contrast Level: Select a display contrast level
from 1(low) to 9 (high) or choose the “default”
(5).
Online/Device
Setup/Review
Displays the current values for all of the pertinent
operating parameters listed in this table, plus
revision information. Parameters are not
configurable in this menu.
Table 12 Tamper Reporting Logic Implementation with Write Protect
Write Protect
Jumper Status
Write Protect
Software Status
Tamper
Reporting Status
Tamper Alert
Posted?
Configuration
Change Allowed?
ON
ON
ON
YES
NO
ON
OFF
ON
YES
NO
OFF
ON
ON
YES
NO
OFF
OFF
ON
YES
YES
ON
ON
OFF
NO
NO
ON
OFF
OFF
NO
NO
OFF
ON
OFF
NO
NO
OFF
OFF
OFF
NO
YES
The sections below give some examples as to how to edit the configuration parameters and execute
Methods.

Table of Contents

Related product manuals