LOAD[:CONFig]
Command :MMEMory:LOAD[:CONfig] <name>[,<msus>][,<module>]
The LOAD command loads a configuration file from the disk into the
modules, software options, or the system. The <name> parameter specifies
the filename from the disk. The optional <module> parameter specifies
which module(s) to load the file into. The accepted values are –2 through
10. Not specifying the <module> parameter is equivalent to performing a
’LOAD ALL’ from the front panel which loads the appropriate file for both the
system and the modules, and any software option.
<name> A string of up to 10 alphanumeric characters for LIF in the following form:
NNNNNNNNNN
or
A string of up to 64 alphanumeric characters for DOS in one of the following
forms:
NNNNNNNN.NNN when the file resides in the current directory or
\NAME_DIR\FILENAME when it does not reside in the current directory
<msus> Mass Storage Unit specifier. INTernal0 for the hard disk drive and
INTernal1 for the flexible disk drive.
<module> An integer, –2 through 5 for the HP 16500C alone. –2 through 10 with the
HP 16501A connected.
Example OUTPUT XXX;":MMEMORY:LOAD:CONFIG ’FILE "
OUTPUT XXX;":MMEMORY:LOADFILE ’,0"
OUTPUT XXX;":MMEM:LOAD:CONFIG ’FILE A’,INTERNAL0,1"
MMEMory Subsystem
LOAD[:CONFig]
12–15

Table of Contents

Other manuals for HP 16500C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 16500C and is the answer not in the manual?

HP 16500C Specifications

General IconGeneral
BrandHP
Model16500C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals