SKEW<N>
Command :INTermodule:SKEW<N> <setting>
The SKEW command sets the skew value for a module. The <N> index value
is the module number (1 through 5 corresponds to the slot location of the
modules A through E for the HP 16500C alone and 1 through 10 to slot
location of modules A through J when an HP 16501A is connected). The
<setting> parameter is the skew setting (– 1.0 to 1.0) in seconds.
<N> An integer, 1 through 5 for HP 16500C alone. 1 through 10 with the HP
16501A connected.
<setting> A real number from –1.0 to 1.0 seconds
Example OUTPUT XXX;":INTERMODULE:SKEW2 3.0E-9"
Query :INTermodule:SKEW<N>?
The query returns the user defined skew setting.
Returned Format
[INTermodule:SKEW<N>] <setting><NL>
Example OUTPUT XXX;":INTERMODULE:SKEW1?"
INTermodule Subsystem