TTIMe
Query :INTermodule:TTIMe?
The TTIMe query returns five values for the HP 16500C alone or ten with an
HP 16501A connected representing the absolute intermodule trigger time for
all of the modules in the Intermodule configuration. The first value is the
trigger time for the module in slot A, the second value is for the module in
slot B, the third value is for slot C, etc.
The value 9.9E37 is returned when:
The module has not yet been run;
No module is installed in the corresponding slot;
The module in the corresponding slot is not time correlated; or
A time-correlatable module did not trigger.
The trigger times returned by this command have already been offset by the
INTermodule:SKEW values and internal hardware skews (INTermodule:HTIMe).
Returned Format
[:INTermodule:TTIMe] <module>,<module>,<module>,<module>,
<module>[,<module>,<module>,<module>,<module>,<module>]<NL>
<module> Trigger time for module (real number)
Example OUTPUT XXX;":INTERMODULE:TTIME?"