Lockout Command
To lockout the front-panel controls, use the SYSTem command LOCKout.
When this function is on, all controls (except the power switch) are entirely
locked out. Local control can only be restored by sending the :LOCKout
OFF command.
Hint Cycling the power will also restore local control, but this will also reset
certain RS-232-C states. It also resets the Logic Analysis System to the
power-on defaults and purges any acquired data in the acquisition memory of
all the installed modules.
See Also For more information on this command see chapter 11, "SYSTem Subsystem."
Programming Over RS-232-C
Lockout Command
3–10