HP
3325B
Remote
Operation
Command
Set
MOFR;
Modulation
Source
Frequency
Command
The
MOFR
command
sets
the
modulation
source
frequency.
Sending
MOFR
with
no
value
or
units
displays
the
current
frequency
Issuing
MOFR?
causes
the
instrument
to
output
its
current
frequency
^
Notes
Only
two
digits
of
frequency
resolution
are
available.
The
timebase
is
not
locked
to
the
main
signal
or
an
external
reference
input
_
Programming
the
frequency
causes
the
signal
to
turn
off
momentarily.
Instrument
Preset
value:
1000.0
Hz
Command
Availability
_
MOFR
MOFR?
HP
3325B
HP
3325A
Yes
Yes
No
No
Table
2-30.
MOFR
“value”
Restrictions
Given
“units”
Value
Range
Units
Description
0.0-^
10000.0
HZ
Hertz
0.0^10.0
KH
Idio-Hz
0.0
0.01
MH
mega~Hz
Table
2-31.
MOFR?
Response
Format
f
i
HEAD-on
response
HEAD-off
response
M0FR########.###HZ
########.###