Chapter 4 Command Reference
:DISPlay Subsystem
4-38 Programming Guide
:DISPlay[:WINDow]:TEXT:FEED [:]CALCulate2 |
[:]CALCulate3
Sets or queries what data flow is fed into the display.
Choose from the following <data_handle> strings:
[:]CALCulate2should be used to direct any result other than the statistics
to the result display
Specifically this would select one of the following results for the result
display:
– raw measurement (if math is disabled:CALC:MATH:STAT OFF)
– the scaled/offset measurement (if math is enabled:CALC:MATH:STAT
ON)
– the limit graph (if limit testing is enabled with the graphic display
:CALC2:LIM:STAT ON, :CALC2:LIM:DISP GRAPh)
[:]CALCulate3should be used to direct the statistical result
(if statistics are enabled, :CALC3:AVER[:STAT] ON) to the result display;
the particular statistic displayed is determined by :CALC3:AVER:TYPE.
A string is returned: CALC2 or CALC3.
*RST: :CALCulate2
Refer to the section titled How to Program the Counter to Display Results
in Chapter 3 of this guide.