Chapter 4 Command Reference
[:SENSe] Subsystem
4-86 Programming Guide
[:SENSe]:FREQuency:ARM[:STARt]:SOURce IMMediate
|EXTernal
Sets or queries the start arm for Frequency, Frequency Ratio, and Period
measurements.
A sequence of ASCII-encoded bytes: IMM or EXT
*RST: IMMediate
Gate & ExtArm
[:SENSe]:FREQuency:ARM:STOP:DIGits <numeric_value>
Sets or queries the resolution in terms of digits used in arming Frequency, Period,
and Ratio measurements.
3 to 15
Numeric data transferred as ASCII bytes in <NR1> format.
*RST: 4
Only applies when [:SENS]:FREQ:ARM:STOP:SOUR DIG is selected.
Gate & ExtArm
[:SENSe]:FREQuency:ARM:STOP:SLOPe POSitive | NEGative
Sets or queries the slope of the external stop arm signal used in external arming
Frequency, Frequency Ratio, and Period measurements.
A sequence of ASCII-encoded bytes: POS or NEG
*RST: NEGative
Only applies when [:SENS]:FREQ:ARM:STOP:SOUR EXT is selected.
Gate & ExtArm