Error Messages
SCPI Error Messages
704
705
706
709
710
711
712
713
-300, “Device specific error;No Sufficiently Wide Pass Band Was
Found(
704)’
No sufficiently wide YTF pass band was found in the initial phase
of the peaking algorithm.
-300, “Device specific error;The fine-peak phase of the peaking
algorithm
failed(705)”
For unspecified reasons the later “tie peak” phase of the peaking
algorithm failed.
-300, “Device specific error;SAF Tracking
Failure(706)”
The SAF tracking algorithm failed for
unspecihed
reasons.
-300, “Device specific error;Calibration Security is LOCKED.
Unable to Access Cal
Data(709)”
Current Calibration Security system is in LOCKED position.
Calibration data is unable for write access. To UNLOCK the
Calibration Security system, refer to the Service Guide for CPU
board dip switch configuration.
-300, “Device specific error;Flatness Calibration
Failed(719)”
The Flatness calibration algorithm failed for
unspecmed
reasons.
-300, “Device specific error;Flatness Calibration Failed (711)”
The Flatness calibration algorithm failed due to an error in reading
power from the external power meter.
-300, “Device specific
error;Flatness
Calibration Failed
Relinquish Failure(7 12)
))
The Flatness calibration algorithm failed experienced difficulties in
relinquishing control of the external power meter.
-300, “Device specific error;No Tracking With
MMH(713)”
The SAF tracking algorithm is not allowed to run under millimeter
head personality. The
millimeter
head should be disconnected hrst.