Safety and Regulatory Information
Ventilation
Requirements
When installing the instrument in a cabinet, the convection into and out of
the instrument must not be restricted. The ambient temperature (outside
the cabinet) must be less than the maximum operating temperature of the
instrument by 4
“C
for every 100 Watts dissipated in the cabinet. If the
total power dissipated in the cabinet is greater than 800 Watts then forced
convection must be used.
Cleaning
Instructions
The connectors on the sweeper should be cleaned periodically for good
connections. The cabinet should only be cleaned using a damp cloth.