Specifications and Options
General
Environmental
Operating temperature range: 0 to 55 
“C
Type tested to the environmental requirements of: MLT-28800E Class
5.
EMC: Conducted and radiated interference comply with:
EN55011
 
class
 
AKISPR-11
 Class A
EN50082-l-1991
IEC 
801-2/1991
 4 
kV
 CD, 8 
kV
 AD
IED
 801-3/1984  3 V/m (26-500 MHz)
IEC 
801-4/1988
 500 V
Warmup
 
Time
Operation: Requires 30 minutes warm-up time from cold start at 0 to 55
“C.
 Internal temperature equilibrium reached after 2 hour warm-up at
stable ambient temperatures.
Frequency Reference (Option lE5 only): Reference timebase is kept at
operating temperature with the instrument connected to AC power.
Attention!
Instruments disconnected from AC power for more than 24 hours require up to 30 days to achieve
timebase
 aging specification. Instruments disconnected from AC power less than 24 hours require 24
hours to achieve 
timebase
 aging specification.