Operation  Model 8901B 
Summary 
of 
Special finction 
37 
Special Function Description 
Selects  the  Automatic  Cal  Factor mode. This  en- 
ables the Modulation Analyzer  to automatically se- 
lect cal factors based on those entered using Special 
Function 37.3. 
Selects the  Manual Cal Factor Mode. This enables 
the user to override the automatic selection of  cal 
factors. 
Procedure: 
To enter a manual cal factor, select: 
37.1 SPCL, cal factor value, 
YO 
CAL FACTOR (Blue 
Key, MHz) 
~~ 
Displays the status of the cal factor mode 
[O=Automatic. 
1 
=Manual). 
~  ~~~~~ 
Prepares to enter a  reference cal factor 
or 
a fre- 
quencylcal-factor  pair into the appropriate table in 
memory. 
Procedures: 
To enter the reference cal factor, select: 37.3  SPCL, 
cal factor value, 
YO 
CAL FACTOR (Blue Key, MHz). 
To enter each frequencylcal-factor pair, select: 
37.3  SPCL, frequency value  (in MHz), MHz key, cal 
factor value, 
YO 
CAL FACTOR (Blue Key, MHz). (It’s 
unnecessary to wait until the values are displayed be- 
fore continuing with subsequent entries.) 
(See examples in this section.) 
Displays the number of entries in the table. 
Prepares to recall the reference cal factor from the 
table. 
Procedure: 
To display the reference cal factor, select: 
37.5 SPCL, 
Yo 
CAL FACTOR (Blue Key, MHz). 
Prepares to recall the next frequency/cal-factor pair 
from the table. 
Procedure: 
To display the frequency, select: 
37.6  SPCL, DISPLAY FREQ (Blue Key, kHz). 
To display the cal factor, select: 
Yo 
CAL FACTOR (Blue Key, MHz). 
37.6 SPCL sequentially accesses frequency/caC 
factor pairs. 
(See example in this section.) 
Clears the table of cal factors from memory. To enter 
a manual cal factor, use 37.1 SPCL. To  enter a new 
table of calibration factors, use 37.3  SPCL. 
Comments 
0 
This mode is selected upon instrument power-up and 
with Automatic Operation. 
0 
Can be keyed from any Measurement mode. 
0 
If no cal factor  is entered, the Modulation Analyzer 
uses the cal factor  recalled last from the table as the 
manual entry. 
0 
The  manual cal factor  remains in effect when  the 
Modulation Analyzer is retuned. 
0 
The automatic cal factor table is not affected by any 
manual cal factor entries. 
0 
Can be keved onlv after selectina RF POWER. 
0 
Can be keyed from any Measurement mode. 
~~ 
____~~ 
____ 
~~ 
0 
If an error is made 
during 
entry of any value,  select 
CLEAR and immediately re-enter the correct value. 
0 
To replace an existing cal factor entry, enter the new 
cal factor at the frequency previously used 
in 
the entry. 
The  Modulation Analyzer will  automatically save  the 
entry in the correct order in the table. 
0 
To  remove an  existing  cal factor  from the table in 
memory, enter 
0% 
as the cal factor  at the frequency 
previously used in the entry. 
0 
Can be keyed from any Measurement mode. 
0 
The number of table entries displayed also includes 
the reference calibration factor. 
0 
Can be keyed from any Measurement mode. 
0 
This special function configures the Modulation An- 
alyzer to use manual cal factors with the reference cal 
factor from the table as the manual entry. To re-enter the 
Automatic Cal Factor mode, select AUTOMATIC OPER- 
ATION 
or 
37.0  SPCL. 
0 
Can be keved onlv after selectina RF POWER. 
0 
The table in memory is accessed sequentially (the 
lowest frequency accessed first). To access the lowest 
frequency in the table, select 37.5 SPCL before select- 
ing 37.6  SPCL. 
0 
This special function configures the Modulation An- 
alyzer to use manual cal factors with the cal factor re- 
called last from the table as the manual entry. To re- 
enter the Automatic Cal Factor mode, select AUTO- 
MATIC OPERATION, then reselect 
RF 
POWER. 
0 
An attempt to read beyond the end of the table re- 
sults in Error 26. 
0 
Displays Error 15 
if 
automatic cal factors were being 
used. 
0 
Can be keved from anv Measurement mode. 
3-74 
RF 
Power