Model 8901B  Operation 
When CALIBRATE is selected, the power reference oscillator outputs a calibrated, 
1 
mW signal at 
50 
MHz. When SAVE CAL is selected, the controller equates the output from the power meter to 
1 
mW. 
The calibration procedure allows the instrument to set the power-measurement transfer function 
so 
the 
displayed power correctly represents  the actual power.  This procedure can be visualized  as a process 
that adjusts the intercept  and slope of  the sensor’s transfer function. The ZERO key sets the zero 
crossing  (intercept), and the SAVE  CAL key  sets the gain  (slope).  (See the following figure.)  The 
displayed power is given by the equation: 
Pdi3 
= 
mPact 
+ 
b 
where; 
Pdi3 
is 
the Displayed Power, 
Pact 
is 
the Actual Power, 
b 
is 
the zero crossing, and 
m 
is the slope. 
The calibration routine forces 
b 
= 
0 
and 
m 
= 
1. 
The following example shows the different responses that occur during calibration: 
DISPLAYED 
POWER 
the 
SAVE CAL 
key 
is 
pressed 
(rn=l) 
ACTUAL POWER 
Minor adjustments in the slope are made by the controller, based on the calibration factors, to correct 
for variations in the power sensor’s sensitivity due to frequency and mismatch. 
RF Power 
3-77