Performance Tests
Special
Function
10.1
10.2
10.3
10.4
10.5
Model 8901B
Power
Reading
Actual Maximum
0.00
pW
f0.06
0.0
pw
fO.l
0.000
pW
fl
count
0.00
rnW
fl
count
0.0
rnW
fl
count
10.1
10.2
10.3
10.4
10.5
Serial Prefix
24614
and
above:
0.000
pW
f0.051
0.00
pW
f0.06
0.0
pW
f2
count
0.000
rnW
fl
count
0.00
rnW
fl
count
I
Saecial
I
Power Reading
I
Range
Calibrator
Range
10
pw
100
pw
10
mW
100
mW
-
I
-r-----
I
I
Power Reading Limits
Lower Actual Upper
9.90
pw
10.10
pw
99.4
pw
100.6
pW
9.94
mW
10.06
mW
99.0
mW
101.0
rnW
Function
I
Actual
I
Maximum
I
I
Range-to-Range
Error
5.
On the range calibrator, set the function
to
calibrate.
6.
On the Modulation Analyzer, key in 10.0
SPCL
to
set the power meter
ranging
to
automatic.
Press CALIBRATE then press the blue key then press
SAVE
CAL (again, the CALIBRATE
key). This normalizes the power reading
to
1
mW.
(The CALIBRATE key light should be on.)
7.
On the Modulation Analyzer, key in the Special hctions
listed
in the following table. For each
setting, set the range calibrator
to
the range indicated
in
the table. The reading on the Modulation
Analyzer should be within the limits shown.
Special
Function
10.2
10.4
10.5
4-30
Performance Test 8
I
I
I
rev.13JUL87