Model 8901B Performance Tests
Accuracy within a Range
8.
Set the range calibrator’s range
to
1
mW.
9. On the Modulation Analyzer, key in 10.3
SPCL.
Set the range calibrator’s range
to
100
pW.
The
Modulation Analyzer’s display should read between 0.099
and
0.101
mW
(0.100
mW
f0.5%
fl
digit, then rounded
to
the nearest 0.001 mW).
Accuracy within a Range: 0.099
0.101 mW