Performance 
Tests 
Model 
8901B 
nble 
4-1. 
Performance Test Record 
(3 
of 
5) 
- 
lest 
No. 
5 
- 
6 
- 
7 
- 
,-42 
lest Description 
~  ~  ~~~  ~  ~~ 
AUDIO  MEASUREMENTS PERFORMANCE TEST 
Audio 
RMS 
Accuracy 
50 
Hz Rate 
500 Hz Rate 
5 kHz Rate 
40 kHz Rate 
40  kHz Rate 
5 
kHz Rate 
500 
Hz Rate 
50 
Hz Rate 
Audio Counter Range and Sensitivity 
Rare 
(Hz) 
RF 
Level 
(dBm, 
V) 
20  +16.5,3 
20  -13.0,  0.1 
1 
000 
-13.0,  0.1 
1 000  +16.5,  3 
250 000  +16.5,  3 
250 000  -13.0,  0.1 
Residual Noise and Distortion 
400  Hz Input at 
100 
mV 
400 Hz Input at 
3V 
1 
kHz Input at 
3V 
1 
kHz Input at 
100 
mV 
Distortion Accuracy 
1 
kHz Rate; 
2nd 
harmonic 
1 
kHz Rate; 
3rd 
harmonic 
1 
kHz Rate; 4th harmonic 
1 
kHz Rate; 5th harmonic 
400 Hz Rate; 2nd harmonic 
400 Hz Rate; 
3rd 
harmonic 
400 Hz Rate; 4th harmonic 
400  Hz Rate; 5th harmonic 
POWER REFERENCE PERFORMANCE TEST 
(USING REFERENCE COMPARISON) 
RF 
Power Reference Accuracy 
POWER REFERENCE PERFORMANCE TEST 
(USING DC SUBSTITUTION) 
RF 
Power Reference Power 
Minimum 
2.88 
Vrms 
2.88 
Vrms 
2.88 
Vrms 
2.88 
Vrms 
96 
mVrms 
96 
mVrms 
96 
mVrms 
96 
mVrms 
3.01 
998 
kHz 
3.01 998 
kHz 
3.99997 
kHz 
3.99997 
kHz 
249.997 
kHz 
249.997 
kHz 
0.89% 
0.89% 
0.89% 
0.89% 
0.89% 
0.89% 
0.89% 
0.89% 
98.1% 
REL 
0.988 
mW 
Results 
Actual 
Maximum 
3.12 
Vrms 
3.12 
Vrms 
3.12 
Vrms 
3.12 
Vrms 
104 
mVrms 
104 
mVrms 
104 
mVrms 
104 
mVrms 
0.02002 
kHz 
0.02002 
kHz 
1.00003 
kHz 
1 .OW03 
kHz 
250.003 
kHz 
250.003 
kHz 
0.3% 
0.3% 
0.3% 
0.3% 
1.12% 
1.12% 
1.12% 
1.1 
2% 
1.12% 
1.1 2% 
1.12% 
1 -12% 
101.9% 
REL 
1.012 
mW 
rev.17AUG87