6
Safety
1. Safety
We recommend that exposure to ultrasound should be kept As Low As
Reasonably Achievable - (ALARA guidelines). This is considered to be
good practice and should be observed at all times.
Team 3 provides just one indicator of fetal condition. This should be
assessed as part of an holistic approach to obstetric care together
with other factors. A complete assessment must be made before taking
appropriate action. If there is any doubt concerning the accuracy of
any measurement, an alternative method should be used.
Symbols
General Warning
Attention, consult this manual. Refer to safety section.
Refer to Instructions
for Use
Attention, consult accompanying
documents / Instructions for Use
1.1 Warnings
Do not use in the presence of fl ammable gases or in oxygen rich
environments.
Do not sterilise the product or its accessories. The product will be
damaged, and there is a risk of patient and user harm.
Do not immerse Team 3 in liquid. Ultrasound and Toco transducers are
IPX7 rated. Team 3 is not intended for use in water birth situations.
Do not use in the sterile fi eld unless additional barrier precautions are
taken.
Use only recommended accessories listed in this manual.
Do not dispose of batteries in fi re as this can cause them to explode.
Do not use with defi brillators.
Do not use with electrosurgical devices.
Team 3 can be isolated from the AC mains supply by removing the IEC
mains inlet connector. Ensure that this is fully accessible at all times.
Team 3 is a Class 1 product that relies for safety on its protective earth.
Ensure it is connected to a suitably earthed AC mains supply.