97
Specifi cations
Maternal oximetry
Sensor types BCI (Smiths Industries) 3444 Finger Sensor
NELLCOR DS100A
Or as detailed in the current Team 3 Accessories &
Consumables Brochure
Saturation range 0-99% SpO2
Saturation
accuracy
±1SD of normal distribution, within ranges:
70–99% ±2 digits
0–69% unspecifi ed
Pulse rate range 30-240 bpm
Pulse rate
accuracy
±2 bpm or ±2%, whichever is greater
Record/display On-screen display and printed record of:
Maternal % SpO2
Heart rate
Safety Type B protection
Alarms Low saturation: 85-99% SpO2
Signal loss: Pulse or ECG
Alarm Sound
Pressure Levels
at 1m
Minimum: 53db(A)
Default 75db(A)
Maximum 92db(A)
Standards ISO9919 : 2005
17.5 Printer
Print head 128mm thick fi lm
Resolution 8 dots per mm
Printer speeds 1, 2, or 3cm per minute (user selectable)
10 cm per minute fast forward
Paper Plain thermal paper, z-fold, 45m length
FHR scales 30–240 bpm or 50–210 bpm (user selectable)
Annotation ● Hospital name, time, date, paper speed, monitoring
modes, signal loss
● Mother’s name and ID number (optional)