74
Dawes-Redman Antepartum Analysis
While reviewing a CTG with analysis results
• If the Criteria are Met, the Dawes-Redman button will be Green.
• If the Criteria are Not Met, the Dawes-Redman button will be Cyan.
• If the Green Dawes-Redman button is pressed, Team 3 will display the
results from the last analysis calculation.
• If the Cyan Dawes-Redman button is pressed, the fi nal results of the
analysis will be displayed.
Stopping Dawes-Redman analysis
Stop recording or printing to stop analysis. The CTG will contain the messages
‘Dawes-Redman Criteria Met’, Dawes-Redman CTG analysis invalid' or ‘Dawes-
Redman Criteria Not Met’.
11.6 Dawes-Redman Analysis report
When the analysis is stopped, the printer produces a report of the analysis
results at the end of the trace. The report shows:
• Values for the calculated parameters
• When the Dawes/Redman criteria were fi rst met
• Whether the Dawes/Redman criteria were met at the time the analysis was
stopped
• Abnormalities
CARE result for
Started at:4
Stopped at:
Dawes-Redman criteria met at
Signal Loss
Fetal Movements per hour
Basal Heart Rate (bpm)
Accelerations
Decelerations > 20 lost beats
High Episodes (Min)
Low Episodes (Min)
Short Term Variatio (ms)
Dawes-Redman analysis is not valid during labour.
This is NOT A DIAGNOSIS.
FHR1
11.08 AM
12.08 PM (60 mins)
60 minutes
0.0%
0.0
140
8
8
14
0
6.4 (2.2 bpm)
Team 3 shows the last calculated
values for
• Signal Loss
• Fetal movements per hour
• Basal heart rate
• Accelerations
• Decelerations
• High episodes
• Low episodes
• Short term variation