72 Digital Refractor CDR-9000
he result of Stereo & Minute Stereo Tests when the tests are executed. Y
ES or NO can be chosen. Default value is YES.
4
Input Coincidence: specifies whether to display the function keys for input
ting the result of Aniseikonia Test when the test is executed. YES or NO c
an be chosen. Default value is YES.
5
Fog: Balance: specifies the amount of fogging power to set automatically
when one of the Binocular Balance tests is executed. Up to +1.00 D can
be set by the step of +0.125D. This value also serves as the alternative v
alue for the amount of fog when the ‘+ Fog’ or the ‘- Fog’ function key
is executed in the SPH adjustment mode. Default value is +0.00D.
6
Fog: Red/Green: specifies the amount of fogging power to set automatical
ly when the Red/Green Test or a test with Red/Green filter is executed. U
p to +1.00 D can be set by the step of +0.125D. This value also serves a
s the primary value for the amount of fog when the ‘+ Fog’ or the ‘- Fo
g’ function key is executed in the SPH adjustment mode. Default value is
+0.00D
7
Startup Mode: specifies from which mode system will start when turning
on the power or when the system is reset by the execution of [CLEAR] b
utton. Choose one from EMPTY, UA, LM, RK or SUBJ. Default is EMPTY. S
ee chapter 11 for mode description.
8
SUBJ Startup: specifies from which code data will be copied and used wh
en entering the SUBJ mode by the [SUBJ] button. Choose one among LM,
RK and ANY. Default value is ANY, see chapter 11 for mode description.