5.4 Replacement parts
Description Item no.
Maintenance kit, central cleaning brush 9480900
Section 6 Chloride sensor (007254)
6.1 Specifications
Specifications are subject to change without notice.
Specification Details
Range 0 to 18,000 mg/L
Accuracy ± 10% or ± 5 mg/L (the larger value)
Resolution 0.01 mg/L
Parameters Cl
–
or sensor mV
Deployment life of sensor cartridge 1 to 2 years
Life
4
of sensor cartridge 1 to 2 years
Maximum depth 15 m (49 ft)
Operating temperature 0 to 50 °C (32 to 122 °F)
6.2 Product overview
The chloride sensor (007254) is green. Refer to Figure 6.
There is a replaceable sensor cartridge installed in the sensor. Refer to Specifications on page 15 for
the deployment life of the sensor cartridge.
The sensor cartridge has a depth limit of 15 meters (49 feet). Remove the sensor cartridge and install
the sensor cap before the sonde is deployed deeper than 15 m (49 ft) to prevent damage to the
sensor cartridge. Refer to the illustrated steps in Figure 7.
To extend the life of the sensor cartridge, remove the sensor cartridge and install the sensor cap
when the sensor is not in use. Refer to the illustrated steps in Figure 7.
Figure 6 Chloride sensor
1 Sensor 3 Sensor cap
2 Sensor cartridge 4 Glass vial for sensor cartridge storage
4
Deployment life plus the storage life
English 15