7.5 Sensor guidelines
Keep deposits off the sensor for best measurements. Examine the sensor to make sure that there
are no contaminants. Keep the substances shown in Table 5 away from the sensor.
Table 5 Contaminants
Substance Damage
Oil, sediment or biofilms Contaminates the sensor. If the sensor is contaminated, clean the sensor.
7.6 Maintenance
7.6.1 Clean the sensor
N O T I C E
To prevent damage to the sensor, do not use organic solvent solutions (e.g., acetone or methanol) to clean the
sensor. Organic solvents cause damage to the plastic housing and sensor lenses.
Clean the sensor after each deployment to remove unwanted material (e.g., oil, biological growth and
dirt). In addition, clean the sensor before and after calibration.
Adjust the sonde deployment time in accordance with how quickly sensor fouling occurs in the
deployment waters.
1. Clean the sensor with:
Mild detergent
Warm water
Clean, soft toothbrush
Soft cloth and/or cotton swabs
Note: Do not use dishwashing liquid to clean the sensor. Dishwashing liquid and household cleaners with skin
moisturizer can put a film on the sensor.
2. Rinse the sensor with clean water.
3. Soak the sensor in clean water for a minimum of 30 minutes.
4. Examine the optical windows. If unwanted material is seen on the optical windows, use optical
tissue or a cotton swab with soap and water to clean the optical windows.
5. Rinse the sensor with clean water.
6. If the substance cannot be removed, contact technical support.
7.7 Accessories
W A R N I N G
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Description Quantity Item no.
Rhodamine Calibration Dye Solution, 21.33 ± 2.5 w/w% 236 mL (8 oz) 007273
Solid Secondary Standard 1 007726
20 English