EasyManua.ls Logo

IKA IC control - Page 19

IKA IC control
30 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19
2. Limits:
Under the option "External (ext)", the maximum and minimum temperature for external temperature control can be
set�
Under "l Δ T (int - ext) l", the maximum difference between the internal and external temperature can be set�
Under "Power Output", the maximum heating output can be set as a percentage�
3. Temperature sensor
1) Calibration:
The internal and the external temperature measurement can be calibrated and adjusted�
You can select 2-point calibration or 3-point calibration for internal and external measurement�
Calibration proceeding (example: 2-point calibration):
> Tempering
Temperature Sensor
Calibration
Reset Calibration
> Temperature Sensor
Calibration
Internal
External
> Calibration
Internal
2 Point Calibration
3 Point Calibration
> Calibration > Internal
2 Point Calibration
Temperature sensor calibration
at
Step: 1 / 2
5.00 C
OK
> Calibration > Internal
2 Point Calibration
Temperature sensor calibration
at
Step: 1 / 2
30.00 C
OK
> Calibration > Internal
2 Point Calibration
Temperature sensor calibration
at
Step: 1 / 2
30.00 C
OK
> Calibration > Internal
2 Point Calibration
Target Temp: 30.00 C
Actual Temp: 25.00 C
Please wait...
Step: 1 / 2
> Calibration > Internal
2 Point Calibration
Target Temp: 30.00 C
Actual Temp: 30.00 C
Step: 1 / 2
OK
> Calibration > Internal
2 Point Calibration
Step: 1 / 2
Enter the temperature from the
measuring device
5.00 C
OK
> Calibration > Internal
2 Point Calibration
Step: 1 / 2
Enter the temperature from the
measuring device
32.00 C
OK
> Calibration > Internal
2 Point Calibration
Step: 1 / 2
Enter the temperature from the
measuring device
32.00 C
OK
> Calibration > Internal
2 Point Calibration
Sensor calibrated successfully
OK
Fig. 28
2) Reset Calibration:
By “Reset Calibration”, the calibration value for the internal or the external temperature sensor will be deleted�
4. Notification:
1) ON:
This menu option allows you to activate/deactivate the “Notification“ function� A green check mark and a beep in-
forming when the target value (hysteresis) has been reached�
2) Hysteresis:
This menu option allows you to set the hysteresis from 0 to ± 2�0 K
3) Sound:
Infinite: Beep until you press the “Back” button�
Beep: Single beep�

Related product manuals