REMOTE OPERATION
Status Reporting
10_15 LDC-3900 Series 69
CHAPTER 3
Bit Reference
Figure 3.13 An LDC-3900 TEC Module Output Off Register
0- LASER Current Limit 8- N/A
1- LASER Voltage Limit 9- LASER Output Changed to be out of Tolerance
2- TEC Output is Off (Channel 1)
Event
10- TEC Output is Off (Channel 4) Event
3- LASER Power Limit (output on) 11- TEC High Temperature Limit (Channel 1)Condition
4- N/A 12- Hardware Error
5- TEC Output is Off (Channel 2)
Event
13- TEC High Temperature Limit (Channel 2) Condition
6- TEC Output is Off (Channel 3)
Event
14- TEC High Temperature Limit (Channel 3) Condition
7- N/A 15- TEC High Temperature Limit (Channel 4) Condition
76543210
76
54
3
2
10
15
14
13 12
11 10
8
9
101112131415
9
8
Logical
OR
Turn Output Off
TEC Output Off
Register
TEC Output Off
Enable Register
TEC:ENABle:OUTOFF <nrf>
TEC:ENABle:OUTOFF?
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&