IMO GmbH Gässlesweg 6 D-75334 Straubenhardt Tel.: 070 82 491 80 Fax: 604 36 www.imo-ag.com info@imo-ag.com
29
GmbH
Message Causes Solution
T2
Warm-up time of water
is overshot
- Heating water defect
- Temperature survey water is incorrect
- Take out washload. Start program again
without washload.
If the error repeats, please contact us.
T3
Warm-up time of air
overshot.
- Heating air defect
- Temperature survey is incorrect
- Take out washload. Start program again
without washload.
- If the error repeats, please contact us.
T4
Pumping off time
overshot
- Switch off machine
- Check waste water hose and spare
cotton filter, change them if required
- Start program again
- If the error repeats, please contact us
T5
Maximum drying time
overshot.
- Air discharge openings by washload
covered.
- Check location of washload
Close the door - Front flap is not completely closed - Close front flap
Program requires spray
arm control
- A program foaming or defoaming
cleaner is selected, although spray
-arm control is switched off.
- Switch on spray arm control or select
another program
- Waste water hose plugged
- Filter candle used up
- Leach pump defect