IDS440 Users Manual version 4.2 Parameter Functions
50
Function 74 Configure Timers
The display prompts “tr” when function 74 is activated. Enter the timer number to be
modified (1 to 5) then press the ENTER key. The first parameter (L) is the time length in
tenths of a second. The second parameter (Fn) is the function number to be executed
when the timer times out.
Parameter 75 Enable Peak Detect
Enter a 1 to enable peak detect, enter a 0 to disable peak detect.
Parameter 76 Debug mode for Event monitors and Scale Basic
The Debug mode is used to ‘debug’ event monitors and scale basic functions. Enter a 1
to activate debug mode, enter a 0 to disable debug mode. When debug mode is enabled
Scale Basic functions ‘single step’. The subroutine stack level and the Function number
are displayed when entering a function. Example: “0 F12”. Press the ENTER key to
begin execution of the function. The instruction address and instruction code are
displayed. Example: “21 255”. Continue to press the ENTER key to step thru the
program. NOTE: the first step displayed for all scale basic functions is a GOTO (244).
Function 78 Initialize ID memory
Use function 78 to initialize id memory after it has been corrupted. The ID memory
becomes corrupted when power is lost (when battery backup is disabled) or when the
battery is allowed to discharge below 2 volts.
Parameter 79 Enable Multi-point Linearization
Enter a to 1 to enable multi-point Linearization, enter a to 0 to disable multi-point
Linearization. The IDS 440 uses the calibration factors set in function 80 below when
multi-point Linearization is enabled.