10
tira85e1-a (2014-10) PxG55x.cp
3 Operation
Via the EtherCAT interface, the following and further data are exchanged in the
standardized EtherCAT protocol:
Pressure reading
Pressure unit (Torr, mbar, Pa)
Zero adjustment
Status and error messages
Status of the switching functions
Set Trip Point for switching functions
Caution
Caution: data transmission errors
Any attempt to simultaneously operate the gauge via the RS232C
Serial Interface and EtherCAT interface or the diagnostic port may
result in incorrect data and data transmission errors.
Therefore, it is inadmissible to simultaneously operate the gauge via
the RS232C Serial Interface and EtherCAT interface or the diagnostic
port.
5
6
4
7
3
9
2
8
1
10
Position Label Function
1 IN EtherCAT IN connector
2 ERR Error LED
3 x100 Address Switch × 100, hexadecimal
4 LA Link activity EtherCAT IN
5 x10 Address Switch × 10, hexadecimal
6 x1 Address Switch × 1, hexadecimal
7 RUN EtherCAT Status LED
8 LA Link activity EtherCAT OUT
9 OUT EtherCAT OUT connector
10 PWR Sensor cable connector
(Power, analog I/O, RS232C I/O and Relay contacts)