18
tira85e1-a (2014-10) PxG55x.cp
Indicates whether the Value parameter contains a value in over range.
Reading Valid
0 No Overrange Exceeded
1 Overrange Exceeded
Indicates whether the Value parameter contains a value in under range.
Reading Valid
0 No Underrange Exceeded
1 Underrange Exceeded
Index SI DataType NV Access PM Name
6010 0x0E BOOL RO tx TxPdoState
0x11 REAL RO tx Sensor Value
Is set if the device is not in Safe State (value (I 0x6nn0, SI 0x11) = valid)
TxPdoState
0
1
Invalid
Valid
The corrected, converted, calibrated final analog input value of the sensor.
Index SI DataType NV Access PM Name
6012 0x01 BOOL RO tx Reading Valid
0x02 BOOL RO tx Overrange Exceeded
0x03 BOOL RO tx Underrange Exceeded
Indicates whether the Value parameter contains a valid value within the specified
accuracy or not.
Reading Valid
0 Invalid
1 Valid
Indicates whether the Value parameter contains a value in over range.
Reading Valid
0 No Overrange Exceeded
1 Overrange Exceeded
Indicates whether the Value parameter contains a value in under range.
Reading Valid
0 No Underrange Exceeded
1 Underrange Exceeded
Subindex 0x02
Subindex 0x03
4.4.3 Input Common
Subindex 0x0E
Subindex 0x11
4.4.4 Input Piezo
Subindex 0x01
Subindex 0x02
Subindex 0x03