EasyManua.ls Logo

Inficon PCG550 - Physical, Environmental, and Dimension Specifications

Inficon PCG550
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16
tira85e1-a (2014-10) PxG55x.cp
Index SI DataType NV Access PM Name
1C00 0x01
0x02
0x03
0x04
BYTE RW Sync Manager Type
1C12 /
1C13
0x01
0x02
0x03
0x04
UINT RW Sync Manager PDO
Assignment
1C32 /
1C33
0x01
-
0x20
RW Sync Manager Parameter
The manufacturer-specific profile objects contain the manufacturer's model number
and device configuration data, status and diagnostic data. The objects are
described in the following tables.
PCG gauges with or without ATM: The "Manufacturer Configuration Module 1"
is assigned to the CDG module.
PSG gauges: The "Manufacturer Configuration Module 1" is assigned to the
Pirani module.
Index SI DataType NV Access PM Name
4000 0x01 BYTE RW Safe State
0x02 REAL RW Safe Value
Specifies the behavior for the value for States other than Valid.
Safe State
0
1
2
3
Zero
Full Scale
Hold last value
Use safe value
Safe Value: The value to be used for Safe State = Safe Value.
PCG gauges with ATM: The "Manufacturer Configuration Module 2" is assigned
to the ATM module.
PCG gauges without ATM: The "Manufacturer Configuration Module 2" is
assigned to the Pirani module.
Index SI DataType NV Access PM Name
4010 0x01 BYTE RW Safe State
0x02 REAL RW Safe Value
Specifies the behavior for the Value for states other than Valid.
Safe State
0
1
2
3
Zero
Full Scale
Hold last value
Use safe value
Sync Manager
4.3 Manufacturer-specific
Profile Objects
(0x2000...0x5FFF)
4.3.1 Manufacturer
Configuration Module 1
Subindex 0x01
Subindex 0x02
4.3.2 Manufacturer
Configuration Module 2
(PCG with or without
ATM only)
Subindex 0x01