tina96e1-b (2016-12) VGC501_VGC502_VGC503.om 41
The measurement value filter permits a better evaluation of unstable or disturbed
measuring signals.
The measurement value filter does not affect the analog output
( 22).
Value
FILTER I
FILTER OFF I No measurement value filter
FILTER FAST I Fast:
The VGC50x responds quickly to fluctua-
tions in the measurement value. As a result,
it will respond faster to interference in the
measured values.
FILTER NORMAL I Normal (factory setting):
Good relationship between response and
sensitivity of the display and the switching
function to changes in the measured values.
FILTER SLOW I Slow:
The VGC50x does not respond to small
changes in measured values. As a result, it
will respond more slowly to changes in the
measured values.
Pressure p
Time t
Pressure p
Time t
Pressure p
Time t