44 tina96e1-b (2016-12) VGC501_VGC502_VGC503.om
The correction factor GAS allows
the measured value to be calibrated for the preset gases N
2
, Ar, H
2
, He, Ne, Kr
and Xe, or
manual input of the correction factor for other gases (COR).
Characteristic curves in [1] … [16].
This parameter is not available for the unit of measurement: Volt.
Available for the following gauges:
Only for pressures
Pirani
(PSG) <1 mbar
Pirani / Capacitance (PCG) <1 mbar
Cold cathode
(PEG, MAG)
Cold cathode / Pirani
(MPG) <1×10
-3
mbar
Hot ionization / Pirani
(BPG) <1×10
-3
mbar
Hot ionization / Pirani (HPG)
Capacitance (CDG)
Hot ionization / Pirani / Capacitance
(BCG) <1×10
-3
mbar
Value
GAS I
GAS N2 I Gas: nitrogen / air (factory setting)
GAS AR I Gas: argon
GAS H2 I Gas: hydrogen
GAS HE I Gas: helium
GAS NE I Gas: neon
GAS KR I Gas: krypton
GAS XE I Gas: xenon
GAS COR I Calibration factor for other gases by
manually entering parameter COR
The calibration factor COR allows the measured value to be calibrated for other
gases ( characteristic curve in [1] … [16]). This parameter is effective in the
entire measurement range of the gauge.
Precondition: Parameter "GAS COR" is set (except capacitance gauges).
This parameter is not available with the measurement unit: Volt.
Available for the following gauges:
Pirani
(PSG)
Pirani / Capacitance (PCG)
Cold cathode
(PEG, MAG)
Cold cathode / Pirani
(MPG)
Hot ionization / Pirani
(BPG, HPG)
Capacitance (CDG)
Hot ionization / Pirani / Capacitance
(BCG)
Value
COR I
e.g. COR 1.00 I No correction
e.g. COR 1.53 I Measurement value corrected by a
factor of 0.10 … 10.00
Correction factor GAS
Calibration factor COR