tina96e1-b (2016-12) VGC501_VGC502_VGC503.om 45
Activating / deactivating the high vacuum measurement circuit (→ also [31]).
Available for the following gauges:
Pirani
(PSG)
Pirani / Capacitance (PCG)
Cold cathode
(PEG, MAG)
Cold cathode / Pirani
(MPG)
Hot ionization / Pirani
(BPG, HPG)
Capacitance (CDG)
Hot ionization / Pirani / Capacitance
(BCG)
Value
HV-CTRL I
HV-CTRL ON I High vacuum measurement circuit
activated
HV-CTRL OFF I High vacuum measurement circuit
deactivated
Switching the emission on and off.
Available for the following gauges:
Pirani
(PSG)
Pirani / Capacitance (PCG)
Cold cathode
(PEG, MAG)
Cold cathode / Pirani
(MPG)
Hot ionization / Pirani
(BPG402 only)
Capacitance (CDG)
Hot ionization / Pirani / Capacitance
(BCG)
Value
EMISSION I
EMISSION AUTO I The emission is switched on and off
automatically by the gauge
EMISSION HAND I The emission is switched on and off
by the user
The symbol lid solid, if the emission is switched on.
Means of selection.
Available for the following gauges:
Pirani
(PSG)
Pirani / Capacitance (PCG)
Cold cathode
(PEG, MAG)
Cold cathode / Pirani
(MPG)
Hot ionization / Pirani
(BPG402 only)
Capacitance (CDG)
Hot ionization / Pirani / Capacitance
(BCG)
Value
FILAMENT I
FILAMENT AUTO I The gauge automatically alternates
between the filaments
FILAMENT FIL 1 I Filament 1 active
FILAMENT FIL 2 I Filament 2 active
Turning the gauge on / off
Emission
Filament