78 tina96e1-b (2016-12) VGC501_VGC502_VGC503.om
Transmit: DCD [,a,a,a] <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,a,a <CR><LF>
Description
a Resolution a =
0 –> AUTO (default)
1 –> One digit
2 –> Two digits
3 –> Three digits
4 –> Four digits
When the PrE (→ 90) is ON and the pressure is in the range p<1.0E-4 mbar the
display resolution of the PSG- and PCG gauges is reduced by one decimal digit.
Transmit: DGS [,a,b,c] <CR>[<LF>]
Description
a Degas gauge 1, a =
0 –> Degas off (default)
1 –> Degas on (3 minutes)
b Degas gauge 2
c Degas gauge 3
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,b,c <CR><LF>
Description
a Degas status gauge 1
b Degas status gauge 2
c Degas status gauge 3
Transmit: EUM [,a,b,c] <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,b,c <CR><LF>
Description
a Emission for measurement channel 1, a =
0 –> Manually
1 –> Automatic (default)
b Emission for measurement channel 2
c Emission for measurement channel 3
DCD - Display resolution
DGS - Degas
EUM - Emission user mode