tina96e1-b (2016-12) VGC501_VGC502_VGC503.om 79
Transmit: FIL [,a,b,c] <CR>[<LF>]
Description
a Filter gauge 1, a =
0 –> Filter off
1 –> Fast
2 –> Normal
3 –> Slow
b Filter gauge 2
c Filter gauge 3
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,b,c <CR><LF>
Description
a Filter time constant gauge 1
b Filter time constant gauge 2
c Filter time constant gauge 3
Transmit: FUM [,a,b,c] <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,b,c <CR><LF>
Description
a Filament for measurement channel 1, a =
0 –> Automatic (default)
1 –> Filament 1
2 –> Filament 2
b Filament for measurement channel 2
c Filament for measurement channel 3
FIL - Measurement value filter
FUM - Filament user mode
BPG402