80 tina96e1-b (2016-12) VGC501_VGC502_VGC503.om
The full scale value of the measurement range (Full Scale) of linear
analog gauges has to be defined by the user. The full scale value of
linear digital gauges and logarithmic gauges is automatically recognized.
Transmit: FSR [,a,b,c] <CR>[<LF>]
Description
a Full scale value gauge 1, a =
0 –> 0.01 mbar
1 –> 0.01 Torr
2 –> 0.02 mbar
3 –> 0.02 Torr
4 –> 0.05 mbar
5 –> 0.05 Torr
6 –> 0.10 mbar
7 –> 0.10 Torr
8 –> 0.25 mbar
9 –> 0.25 Torr
10 –> 0.50 mbar
11 –> 0.50 Torr
12 –> 1 mbar
13 –> 1 Torr
14 –> 2 mbar
15 –> 2 Torr
16 –> 5 mbar
17 –> 5 Torr
18 –> 10 mbar
19 –> 10 Torr
20 –> 20 mbar
21 –> 20 Torr
22 –> 50 mbar
23 –> 50 Torr
24 –> 100 mbar
25 –> 100 Torr
26 –> 200 mbar
27 –> 200 Torr
28 –> 500 mbar
29 –> 500 Torr
30 –> 1000 mbar
31 –> 1100 mbar
32 –> 1000 Torr
33 –> 2 bar
34 –> 5 bar
35 –> 10 bar
36 –> 50 bar
b Full scale value gauge 2
c Full scale value gauge 3
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,b,c <CR><LF>
Description
a Full scale value gauge 1
b Full scale value gauge 2
c Full scale value gauge 3
FSR - Measurement range
(linear gauges)