EasyManua.ls Logo

Inficon VGC501 - Page 81

Inficon VGC501
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
tina96e1-b (2016-12) VGC501_VGC502_VGC503.om 81
Transmit: GAS [,a,b,c] <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,b,c <CR><LF>
Description
a Gas type correction measurement channel 1, a =
0 –> nitrogen / air (default)
1 –> Argon
2 –> Hydrogen
3 –> Helium
4 –> Neon
5 –> Krypton
6 –> Xenon
7 –> Other gases
b
Gas type correction measurement channel 2
c
Gas type correction measurement channel 3
Transmit: HVC [,a,b,c] <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,b,c <CR><LF>
Description
a Gauge 1, a =
0 –> Off
1 –> On
b Gauge 2
c Gauge 3
Transmit: ITR <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: aa,aa,aa,aa,aa,aa,aa,aa bb,bb,bb,bb,bb,bb,bb,bb
cc,cc,cc,cc,cc,cc,cc,cc <CR><LF>
Description
aa,aa,aa,aa,aa,aa,aa,aa Data string gauge 1
(byte 0 … 7 in hex format)
bb,bb,bb,bb,bb,bb,bb,bb
Data string gauge 2
(byte 0 … 7 in hex format)
cc,cc,cc,cc,cc,cc,cc,cc
Data string gauge 3
(byte 0 … 7 in hex format)
GAS - Gas type correction
HVC - HV control, EMI on / off
ITR - Data output BPG, HPG,
BCG, CDGxxxD

Related product manuals